DEM AudioLinFIR Documentation
LinFIR

Measurement Calibration


On this page
Measurement Calibration Settings Measurement Calibration Settings

Correct measurements for the measurement chain itself, in two independent parts:

  • Microphone: the measurement microphone's frequency response and sensitivity, from its calibration file
  • Soundcard: the audio interface's own response, measured through a loopback cable

Each part keeps a library of calibrations in application settings, with a default for newly created drivers. Each driver then selects its own microphone and soundcard calibration (or none) in its IR window - see IR Management → Measurement Calibration.

These correct the measurement chain's frequency response. To read the measurements in dB SPL, the chain's absolute sensitivity is measured per driver, with a calibrator or a sound level meter, in the same Measurement Calibration section of the IR window - see IR Management → SPL Calibration.

Microphone

Manage calibration files for measurement microphones and define default calibrations for new drivers.

Overview

Microphone calibration compensates for the frequency response and sensitivity of your measurement microphone. LinFIR supports:

  • Multiple calibration files stored in application settings
  • Per-driver calibration selection in the IR Management window (see IR Management)
  • Default calibration for newly created drivers
  • Automatic application to new projects when a default is set
  • Minimum-phase correction computed in real-time from calibration data

Calibration scope:

  • Calibration data is embedded in project files (.lnf) for portability
  • Calibrations from projects are also stored in application settings for reuse
  • Each driver can use a different calibration file or none at all

Calibration File List

The main section displays all imported calibration files in a scrollable grid:

ColumnContent
NameCalibration name, from the imported file (hover for the full name when truncated)
SensitivityMicrophone sensitivity factor in dB
PointsNumber of frequency/magnitude pairs in the file
Serial NumberMicrophone serial number (if present in file), — otherwise
In UseYes if used in current project (hover to see driver names), — otherwise
PlotShows the calibration's frequency response
DeleteRemoves the calibration

Automatic driver updates:

  • When you delete a calibration file that's currently assigned to one or more drivers, those drivers are automatically reset to "None"
  • A warning toast will notify you which drivers were affected
  • This prevents broken references and ensures project stability

Default Calibration

Select a calibration file to automatically apply to newly created drivers.

Dropdown selector:

  • Shows "None" when no default is set
  • Lists all imported calibration files by name
  • Changes apply immediately to new drivers and projects

When the default is applied:

  • Manual driver creation: When you add a new driver via the Drivers toolbar
  • New Loudspeaker Design project: All initial drivers receive the default calibration
  • New Room Calibration project: The initial measurement receives the default calibration

Note: Changing the default does not affect existing drivers in the current project or previously saved projects. Only new drivers created after the change receive the new default.

Import Calibration File

Click "Import Calibration File" to load a microphone calibration file.

Supported file format:

  • Plain text files with two or three columns: Frequency (Hz), Magnitude (dB), and optionally Phase (degrees)
  • Whitespace or comma-separated
  • Lines starting with non-numeric characters are treated as comments/headers
  • Optional header line with sensitivity factor and serial number

Note on phase column: If a third column (phase) is present, it is intentionally ignored. Phase data exported by calibration software is typically smoothed, wrapped, or otherwise processed in ways that make it unsuitable for direct use in transfer function reconstruction. Since measurement microphones exhibit predominantly minimum-phase behavior, reconstructing the phase from the magnitude alone (via Hilbert transform) is both accurate and more robust than relying on the exported phase values.

Example file format:

# Sens Factor = -2.5dB, SERNO: 41486651
# Frequency (Hz)  Magnitude (dB)
20.0              -0.5
50.0              -0.3
100.0             -0.1
200.0              0.0
500.0              0.1
1000.0             0.0
2000.0            -0.2
5000.0            -0.5
10000.0           -1.0
20000.0           -2.0

Sensitivity factor and serial number:

  • Sensitivity factor (in dB) is automatically extracted from the file header if present (line containing "Sens Factor")
  • Serial number is automatically extracted if present (SERNO: field in header)
  • If no sensitivity factor is found in the file, it defaults to 0 dB
  • Serial number is used to differentiate calibrations with identical filenames when sharing projects

Storage: Calibration data (frequency response, sensitivity, and serial number) is embedded in project files (.lnf) to ensure portability:

  • Projects can be shared without requiring recipients to have the calibration files
  • When loading a project, calibrations are automatically added to application settings if not already present
  • If a calibration with the same name exists:
    • With serial number: differentiated using serial number (e.g., filename (SN:7016078))
    • Without serial number: differentiated using numeric suffix (e.g., filename (1), filename (2))
  • You can also manually import calibrations to your settings library for reuse across projects

How Calibration is Applied

LinFIR turns the calibration's magnitude into a minimum-phase correction filter, with the sensitivity as a gain; the phase column of the file, if any, is not used. Calibration Processing explains how and why.

Per-Driver Calibration Selection

Each driver can use a different calibration file (or none). To assign a calibration to a specific driver, use the Microphone dropdown of the Measurement Calibration section in the IR Management window.

See: IR Management → Measurement Calibration

Best Practices

File organization:

  • Use descriptive filenames that include microphone model and serial number
  • Example: UMIK-1_SN123456.txt or Earthworks_M30_SN789.txt

Calibration maintenance:

  • Import calibration files for all your measurement microphones
  • Set your most frequently used microphone as the default

Project workflow:

  • Assign calibrations in the IR Management window after importing measurements
  • Use "None" for measurements taken with already-flat reference microphones
  • Different drivers in the same project can use different calibrations (useful when combining measurements from multiple sessions)

Frequency range:

  • Ensure calibration files cover the full measurement bandwidth
  • LinFIR interpolates between provided points, but extrapolation beyond the calibration range may produce inaccurate results
  • Most measurement microphones provide calibration data from 20 Hz to 20 kHz

Sensitivity factor:

  • Use the value from your microphone's calibration certificate
  • This is typically a negative dB value (e.g., -0.5 dB, -2.4 dB)
  • If unsure, use 0 dB and rely solely on the frequency response correction

Soundcard

Measure your audio interface's own response once, and remove it from every measurement taken through it.

Overview

An audio interface is not perfectly transparent: its converters and their filters shape the magnitude and phase of everything you measure, mostly at the frequency extremes (phase rotation in the low bass, roll-off and phase shift near Nyquist). A soundcard calibration measures that response through an output-to-input loopback cable and corrects it with a regularized (Kirkeby) inverse filter:

  • Magnitude and phase are both corrected, from the measured response itself
  • Timing is preserved: the corrected impulse response stays where it was, so window positions and delays are unaffected
  • Overall level is unchanged: only the response's shape is corrected
  • Correction range follows the sweep the calibration was measured with; outside it, measurements are left untouched
  • Independent of the microphone calibration: a driver can use either, both, or neither

Default Calibration

Select a calibration to automatically assign to newly created drivers. As for microphones, changing the default does not affect existing drivers.

Measure a New Calibration

  1. Connect the Output Channel directly to the Input Channel with a cable - no amplifier, no microphone
  2. Select both channels, then click Start Measurement
  3. When the measurement completes, give it a Name and click Save (or Discard)

The measurement uses a 5 s sweep at the sample rate currently set in Audio Settings.

Sample rate matters. A soundcard calibration only corrects measurements captured at the sample rate it was measured at: an interface's response changes from one sample rate to another, so a calibration is never resampled. Measure one calibration per sample rate you work at. In a driver's IR window, calibrations measured at another sample rate than the driver's measurements are greyed out, and a warning is shown if the selected one doesn't match.

Calibration Library

ColumnContent
NameName given when saving the measurement
LengthImpulse response length, in samples and milliseconds
Sample RateSample rate the calibration was measured at
In UseYes if used in the current project (hover to see driver names)
PlotShows the calibration's frequency response
DeleteRemoves the calibration; drivers using it are reset to "None", with a warning

Storage: like microphone calibrations, soundcard calibrations are embedded in project files and added to application settings when a project is loaded.

How the Correction is Applied

Applied to each measurement before its time window, as a regularized inverse of the loopback response: see Calibration Processing.

Best Practices

  • Measure through the same interface, channels, and sample rate as your driver measurements
  • Measure again after changing the interface, its routing, or the sample rate you work at
  • On a loopback, the interface's own distortion is part of both the measurement and the calibration: interpret the THD of a loopback measurement with care